ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, HANDPIECE; UNIT, PHACOFRAGMENTATION
|
Back to Search Results |
|
Catalog Number ASKU |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
A health professional reported no phacoemulsification power with the handpiece.Procedure details were provided.There was no harm to the patient.Additional information has been requested.
|
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
The company service representative examined the system and the event was not duplicated.The system was then tested and met all product specifications.The phacoemulsification handpiece was not returned for evaluation.The serial number (s/n) was not provided and could not be determined based on the information provided.Therefore, manufacturing information could not be obtained.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
|
|