Physician intended to use a hawkone atherectomy device with a spider fx during treatment of a calcified lesion in the patient¿s proximal right superficial femoral artery (sfa).
Moderate vessel calcification is reported.
Lesion exhibited 70-80% stenosis.
Ifu was followed.
Vessel pre-dilation was performed.
Several passes were made with the hawkone device.
When the physician attempted to insert the device to make the final pass, the physician felt severe resistance during advancement and the hawkone catheter could not be pushed forward due to it becoming entangled in the spider fx device.
The physician then removed the hawkone catheter and spider fx.
The procedure was completed using a non-medtronic (mustang) balloon and inpact pacific drug-coated balloon.
The physician attempted to flush the spider fx and hawkone devices following removal from the patient and observed some plastic debris from the nosecone.
The physician suspects the plastic debris is from the nosecone after several uses of the hawkone, as some part of the material peeled off from the nosecone.
There was no detached components.
The cutter was located inside the housing during device removal from the patient.
No patient injury reported.
When the device was returned to the manufacturing facility, it was noted that the device was damaged.
|