• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LAG SCREW DRILL SLEEVE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LAG SCREW DRILL SLEEVE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674523
Device Problems Loose or Intermittent Connection (1371); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, a lag screw drill sleeve was popping out of drops in the jig.Incident occurred during use with instruments outside the patient.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned device could not confirm the stated failure mode but could confirm that the device has scratches on it.The device shows signs of extensive use.A functional evaluation of the returned device confirmed the stated failure mode.The device was paired with the mating part but could not function, rendering the device inoperable.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW DRILL SLEEVE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11205393
MDR Text Key227919742
Report Number1020279-2021-00517
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00885556363249
UDI-Public00885556363249
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674523
Device Catalogue Number71674523
Device Lot Number19CCT0052
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2021
Date Manufacturer Received04/29/2021
Patient Sequence Number1
-
-