Model Number 25-40-45 |
Device Problem
Gas Output Problem (1266)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/22/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 gas blender system was found to have o2 readings out of tolerance.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H10: in the initial report the product code was wrongly set to dwc.The correct code is dtx and has been aligned accordingly in d section.
|
|
Event Description
|
See initial report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: the affected device was received at the manufacturer site for repair and the reported issue could not be reproduced.The gas blender was cleaned and re-calibrated.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Based on the fact that the device worked within specification, the cause of the reported event is most likely related the hospital gas supply issues.
|
|