Catalog Number 0998-00-3023-53 |
Device Problem
Inability to Auto-Fill (1044)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.The fse checked the iabp for leaks and evacuated condensation.The luer fitting was replaced as a precaution.The fse completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.
|
|
Event Description
|
It was reported that the cs300 intra-aortic balloon pump (iabp) had an autofill failure.It is unknown the circumstances under which the event occurred.It is also unknown if there was patient involvement.However, there was no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, b5, d5, e2, e3, g4, g7, h2, h6 (patient codes), h10.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) had an autofill failure.There was no patient harm and no adverse event was reported.
|
|
Manufacturer Narrative
|
Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 12 month product complaint trend data for the period ( feb-2020 through jan-2021) was reviewed.There were no triggers identified for the review period.
|
|
Event Description
|
N/a.
|
|