• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MAKO ROBOTIC ARM 3.1; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MAKO ROBOTIC ARM 3.1; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 219999
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2021
Event Type  malfunction  
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event Description
Mps reported green not disappearing in cuts.During a tka case, green bone was not disappearing with cuts.Mps shutdown and upon rebooting got a software error.Already spoke with fse.Requested logs.Surgery not completed robotically.
 
Event Description
Mps reported green not disappearing in cuts.During a tka case, green bone was not disappearing with cuts.Mps shutdown and upon rebooting got a software error.Already spoke with fse.Requested logs.Surgery not completed robotically.
 
Manufacturer Narrative
Reported event.An event regarding software error involving a mako tka software was reported.The event was not confirmed.Method & results.-product evaluation and results: review of the case session files was not performed as case session data was not provided.The field service engineer reported: problem reproduced? no.Trouble shooting notes: none.Work performed: removed covers and inspected.Ran transmission, phase check, find friction, and optical compliance, all nominal.Re-installed tha and tka software applications.Robot was in great shape, minor cable tensioning.Work order disposition: system investigation completed successfully as per service manual.All system checks and tests passed, system is ready for use.-clinician review: no medical records were received for review with a clinical consultant.-product history review: a review of device history records shows that rob1083 was inspected on 27 jan 2020 and the quality inspection procedures were completed with no reported discrepancies.-complaint history review: a review of complaints in trackwise related to p/n 219999, robot number: rob1083 shows 0 similar complaints for tka software - software error.Conclusions: the alleged failure mode was not confirmed through onsite inspection carried out by the field service engineer and the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the case session/log data are needed to complete the investigation for determining root cause.Successfully completed all checks, verifications, and calibrations.System is ready for clinical use.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAKO ROBOTIC ARM 3.1
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key11217432
MDR Text Key242411045
Report Number3005985723-2021-00011
Device Sequence Number1
Product Code OLO
UDI-Device Identifier07613327395280
UDI-Public07613327395280
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number219999
Device Catalogue Number219999
Was Device Available for Evaluation? No
Date Manufacturer Received03/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-