Reported event.An event regarding software error involving a mako tka software was reported.The event was not confirmed.Method & results.-product evaluation and results: review of the case session files was not performed as case session data was not provided.The field service engineer reported: problem reproduced? no.Trouble shooting notes: none.Work performed: removed covers and inspected.Ran transmission, phase check, find friction, and optical compliance, all nominal.Re-installed tha and tka software applications.Robot was in great shape, minor cable tensioning.Work order disposition: system investigation completed successfully as per service manual.All system checks and tests passed, system is ready for use.-clinician review: no medical records were received for review with a clinical consultant.-product history review: a review of device history records shows that rob1083 was inspected on 27 jan 2020 and the quality inspection procedures were completed with no reported discrepancies.-complaint history review: a review of complaints in trackwise related to p/n 219999, robot number: rob1083 shows 0 similar complaints for tka software - software error.Conclusions: the alleged failure mode was not confirmed through onsite inspection carried out by the field service engineer and the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the case session/log data are needed to complete the investigation for determining root cause.Successfully completed all checks, verifications, and calibrations.System is ready for clinical use.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|