Model Number 10623 |
Device Problem
Material Deformation (2976)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/28/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that stent damage occurred.A 2.25 x 38mm synergy ii drug eluting stent was selected for use.However, after the device was unpacked, it was noticed that the appearance of the stent was not smooth.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: synergy ous mr 2.25 x 38mm stent delivery system (sds) was returned for analysis.A visual examination of the stent found evidence of damage.The struts on the proximal end lifted and pulled distally.The undamaged crimped stent outer diameter was measured using snap gauge and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
|
Event Description
|
It was reported that stent damage occurred.The 80% stenosed target lesion was located in the proximal end of the left anterior descending artery.A 2.25 x 38mm synergy ii drug eluting stent was selected for use.However, after the device was unpacked, it was noticed that the appearance of the stent was not smooth.The procedure was completed with another of same device.There were no patient complications reported and patient status was stable.
|
|
Search Alerts/Recalls
|