Model Number 3662
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/05/2021
Event Type  Malfunction  
Manufacturer Narrative

The device is included in the neuromodulation implantable pulse generator (ipg) inaccurate elective replacement indicator advisory notice issued by abbott on (b)(6) 2017. The results of the investigation are inconclusive since the device was not returned for analysis. Actions have been taken to prevent reoccurrence.

 
Event Description

It was reported that diagnostics indicated the elective replacement indicator (eri) triggered earlier than intended. The wireless software update was performed clearing the eri message. The device is providing therapy.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROCLAIM 7 ELITE IMPLANTABLE PULSE GENERATOR
Type of DeviceSCS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key11219078
MDR Text Key257807177
Report Number1627487-2021-00358
Device Sequence Number1
Product Code LGW
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberP010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,CONSUM
Reporter Occupation
Remedial Action Other
Type of Report Initial
Report Date 01/22/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received01/22/2021
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device EXPIRATION Date10/02/2019
Device MODEL Number3662
Device Catalogue Number3662
Device LOT Number6145109
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/15/2021
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured10/02/2017
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/09/12/2017/001-C

-
-