BOSTON SCIENTIFIC CORPORATION VISIONIST CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U225 |
Device Problems
Difficult to Insert (1316); Material Integrity Problem (2978)
|
Patient Problems
No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during a box change procedure, this cardiac resynchronization therapy pacemaker (crt-p) was an attempted implant.The physician was unable to connect the right atrial (ra) lead into the ra port.It was not possible to advance the lead three-quarters of the way into the port, therefore the lead tip did not reach the set screw.The lead was able to be inserted and connected onto the right ventricular port.This device was replaced and the second device did not encountered any header issues.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, under-insertion of the lead terminal pin into the device header was suspected to be the cause of the insertion difficulty but this could not be confirmed during laboratory analysis.Note that boston scientific has a vigilant quality monitoring system and we aggressively monitor field performance of all our products.We will continue to monitor for similar events.
|
|
Event Description
|
It was reported that during a box change procedure, this cardiac resynchronization therapy pacemaker (crt-p) was an attempted implant.The physician was unable to connect the right atrial (ra) lead into the ra port.It was not possible to advance the lead three-quarters of the way into the port, therefore the lead tip did not reach the set screw.The lead was able to be inserted and connected onto the right ventricular port.This device was replaced and the second device did not encountered any header issues.No adverse patient effects were reported.
|
|