• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.During the evaluation, the housing contained dust inside and out.A loose video connector latch was found causing an unstable flickering image.The software version was upgraded to the latest.No other issues were found during the device inspection.If additional information is obtained a supplemental report will be filed.
 
Event Description
A user facility sent a video system center to olympus for an annual inspection.No device problems were reported at the time.No patient involvement or injuries were reported.No additional information has been obtained.
 
Manufacturer Narrative
The investigation has been completed.A device history record (dhr) review was performed and found no non-conformances that would cause the reported malfunction.All records indicate that the device was manufactured in accordance with all applicable procedures.The instructions for use (ifu) states: ¿do not apply excessive force to the camera cable of the camera head by bending, stretching or crushing it.Also do not pull a bundle of camera cables, as this may cause internal wire disconnection.Do not connect or disconnect the endoscope connector while this video system center is turned on.Connecting or disconnecting the endoscope while this video system center is on may destroy the ccd.Turn the video system center off before connecting or disconnecting the endoscope.¿.The root cause could not be conclusively determined.Probable causes that could have led to the reported event include that the lock or the pin on the receptacle unit was worn away due to repeated use for a long period and failed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11222612
MDR Text Key228477055
Report Number8010047-2021-01832
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/26/2021
Is This a Reprocessed and Reused Single-Use Device? No