• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VISIONIST X4 CRT-P PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VISIONIST X4 CRT-P PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U228
Device Problems Signal Artifact/Noise (1036); Premature Discharge of Battery (1057); Impedance Problem (2950); High Capture Threshold (3266)
Patient Problems No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that after an atrioventricular node ablation procedure, this device exhibited an increase in capture threshold and unable to measure the leads impedance. There was noise recorded. Additionally, the device battery status showed the remaining capacity at one year remaining. Chest x-ray will be perform. The involved right ventricular (rv) lead is a non-boston scientific product. Additional information was received, stating that the device was explanted. No additional adverse patient effects were reported. At this time, the product has been returned, investigation will be performed and this report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVISIONIST X4 CRT-P
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11227765
MDR Text Key228613758
Report Number2124215-2020-29024
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberU228
Device Catalogue NumberU228
Device Lot Number751928
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 01/25/2021 Patient Sequence Number: 1
-
-