Model Number 283512
Device Problems Material Frayed (1262); Use of Device Problem (1670); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Udi: (b)(4). Investigation summary: the device was received and evaluated at the service center. The reported complaint that the device doesn't turn due to a jammed motor, an o-ring was missing and that the cable was damaged, was confirmed. It was found that the motor did not run constantly as it was corroded. Further, the resistance values of the keypad of the hand control set and the protective earth resistance of the motor cable were out of range and the buttons of the keypad were not functioning, the motor was sticking and also the cable and o-ring were damaged. The motor, motor cable and the hand control set were replaced, the device was modified as per the specifications of the manufacturer, repaired, tested and found to be fully functional. Fluid ingress into the system and contact with the motor is responsible for the corrosion of the motor. The corroded motor has a tendency to stick and not turn, hence is responsible for the issue reported by the customer. The damaged cable and the missing o-ring are most likely a result of user mishandling of the device. However, given the information provided, we cannot determine a definitive root cause for the defective keypad of the hand control set. The service history has been reviewed in lieu of the device history record for this device since it was previously serviced. The device was last serviced on 07/08/2020 and passed all functional testing before being returned to the customer. At this point in time, no corrective action is required, and no further action is warranted. Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the customer in (b)(6) that during an unknown procedure on an unknown date, it was observed that the micro tornado hp w hand control device had a jammed motor, missing o-ring, damaged cable, and did not turn. During in-house engineering evaluation, it was determined that the motor on the device did not run constantly as it was corroded. Another like device was used to complete the procedure without delay. There were no adverse patient consequences reported. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMICRO TORNADO HP W HANDCONTROL
Type of DeviceLINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key11228971
MDR Text Key228653344
Report Number1221934-2021-00238
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/25/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/25/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A

-
-