MEDTRONIC MEXICO NC EUPHORA RX CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number NCEUP3512X
Device Problems Detachment of Device or Device Component (2907); Material Deformation (2976)
Patient Problem Insufficient Information (4580)
Event Date 01/21/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
During a procedure an attempt was made to use one nc euphora ptca balloon catheter to treat a lesion. The device was inspected with no issues noted. Negative prep was performed with no issues. It was reported that the device detached/fractured at the catheter. Excessive force was not used. The device was not kinked and re-straightened during use. Another balloon was used to trap the catheter in the guide and everything was removed with no damage to the vessel. It was indicated that the catheter/delivery system deformed. No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNC EUPHORA RX
Type of DeviceCATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX 22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX 22570
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key11229819
MDR Text Key228663001
Report Number9612164-2021-00321
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberNCEUP3512X
Device Catalogue NumberNCEUP3512X
Device Lot Number220611833
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 01/26/2021 Patient Sequence Number: 1
-
-