Device evaluated by mfr.
: returned product consisted of a quantum maverick balloon catheter.
The shaft, hypotube, tip and balloon were microscopically and visually examined.
There was a kink 67.
7cm distal from the strain relief.
There was a complete separation at 70.
5cm distal of the strain relief.
There was blood in the guidewire lumen.
The balloon was tightly folded.
Inspection of the remainder of the device presented no other damage or irregularities.
|
Reportable based on device analysis completed on 08-jan-2021.
It was reported that shaft kink occurred.
A 2.
75mm x 15mm quantum maverick balloon catheter was selected for use.
However, during unpacking, it was noted that the connector of the balloon and the catheter was kinked.
The procedure was completed with another of the same device.
There were no patent complications and the patient was stable.
However, returned device analysis revealed shaft detached/separated.
|