Model Number 7015
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/10/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr. : returned product consisted of a quantum maverick balloon catheter. The shaft, hypotube, tip and balloon were microscopically and visually examined. There was a kink 67. 7cm distal from the strain relief. There was a complete separation at 70. 5cm distal of the strain relief. There was blood in the guidewire lumen. The balloon was tightly folded. Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
Reportable based on device analysis completed on 08-jan-2021. It was reported that shaft kink occurred. A 2. 75mm x 15mm quantum maverick balloon catheter was selected for use. However, during unpacking, it was noted that the connector of the balloon and the catheter was kinked. The procedure was completed with another of the same device. There were no patent complications and the patient was stable. However, returned device analysis revealed shaft detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameQUANTUM MAVERICK
Type of DeviceCATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11233563
MDR Text Key228870476
Report Number2134265-2021-00820
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P860019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial
Report Date 01/26/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/14/2022
Device Model Number7015
Device Catalogue Number7015
Device Lot Number0023492065
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-