Model Number 4542
Device Problems Fracture (1260); Low impedance (2285)
Patient Problem No Code Available (3191)
Event Date 09/01/2018
Event Type  Injury  
Manufacturer Narrative
This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.(b)(4).
 
Event Description
It was reported that this left ventricular (lv) lead was fractured.The boston scientific technical services (ts) reviewed the report and the lead appears to exhibit out of range (oor) pacing impedance.Moreover, there was a plan for a device changeout and lead revision in a few weeks.Additional information indicated that the left ventricular (lv) lead was surgically abandoned.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
RQ   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11233772
MDR Text Key228833916
Report Number2124215-2021-02228
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410901
UDI-Public00802526410901
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/11/2016
Device Model Number4542
Device Catalogue Number4542
Device Lot Number304507
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/11/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age84 YR
-
-