• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) Back to Search Results
Model Number U128
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Pocket Erosion (2013); No Code Available (3191)
Event Date 01/12/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) system was explanted due to infection and pocket erosion.This crt-p system was succesfully replaced.No additional adverse patient effects were reported.
 
Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed, and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.Patient code 3191 was applied to capture the reportable event of surgery.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) system was explanted due to infection and pocket erosion.This crt-p system was successfully replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE X4 CRT-P
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11234037
MDR Text Key228838395
Report Number2124215-2021-02321
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date05/12/2022
Device Model NumberU128
Device Catalogue NumberU128
Device Lot Number752383
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-