Model Number D153 |
Device Problems
High impedance (1291); Over-Sensing (1438); Defective Device (2588)
|
Patient Problems
Palpitations (2467); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/16/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that impedance on the right atrial (ra) channel of this implantable cardioverter defibrillator was high and out of range.The ra lead is from another manufacturer.Oversensing of signals from the respiratory rate trend (rrt) feature were noted on the ra channel.A technical services (ts) consultant discussed turning rrt off and noted that the reported clinical observations could be due to either a lead or a connection issue.Ts also suggested further system evaluation.The patient was seen in clinic and the noise could not be reproduced.Thresholds were noted to be satisfactory.Ts confirmed this indicated that the issue is likely related to the lead-device connection but also suggested continued monitoring of the lead.Rrt was programmed off to optimize device performance.The patient also reported palpitation symptoms.Ts review of electrograms (egm) from the time of the symptoms indicated that the device seemed to be pacing appropriately, no record of a fast rate was found on the egm.The caller reported they would discuss the ts findings with the physician.No adverse patient effects were reported.This device remains in service.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.Additionally, investigation of the available information determined this device exhibited oversensing of noise generated by the minute ventilation (mv)/respiratory sensor that is related to a high impedance condition.Please see the description for more information regarding the specific circumstances of this event.
|
|
Event Description
|
It was reported that impedance on the right atrial (ra) channel of this implantable cardioverter defibrillator was high and out of range.The ra lead is from another manufacturer.Oversensing of signals from the respiratory rate trend (rrt) feature were noted on the ra channel.A technical services (ts) consultant discussed turning rrt off and noted that the reported clinical observations could be due to either a lead or a connection issue.Ts also suggested further system evaluation.The patient was seen in clinic and the noise could not be reproduced.Thresholds were noted to be satisfactory.Ts confirmed this indicated that the issue is likely related to the lead-device connection but also suggested continued monitoring of the lead.Rrt was programmed off to optimize device performance.The patient also reported palpitation symptoms.Ts review of electrograms (egm) from the time of the symptoms indicated that the device seemed to be pacing appropriately, no record of a fast rate was found on the egm.The caller reported they would discuss the ts findings with the physician.No adverse patient effects were reported.This device remains in service.
|
|
Search Alerts/Recalls
|