Model Number 9500-66
Device Problem Imprecision (1307)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/14/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter occurred. The sensor was inserted into the abdomen on (b)(6) 2021. Data was evaluated and the allegation was confirmed. The probable cause could not be determined via data. The reported glucose values fall within the d zone of the parkes error grid. The data investigation found a difference in values that falls within the c zone of the parkes error grid. No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key11238857
MDR Text Key230039678
Report Number3004753838-2021-20125
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation
Type of Report Initial
Report Date 01/27/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date08/19/2021
Device Model Number9500-66
Device Catalogue NumberSTS-GF-012
Device Lot Number5279491
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-