MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH NCB, SCREW, 5.0, 34 MM; NCB PLATING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH NCB, SCREW, 5.0, 34 MM; NCB PLATING SYSTEM Back to Search Results
Model Number N/A
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Appropriate Term/Code Not Available (3191)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Swelling (2091)
Event Type  Injury  
Manufacturer Narrative
The manufacturer received other source of documents for review.As no lot number was provided, the device history records could not be reviewed.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.(b)(4).
 
Event Description
Patient was implanted on an unknown side and after implantation the patient started experiencing pain, extreme swelling and limited range of motion.Patient gained weight post implantation.Hence, the patient went for second opinion and from the x-rays it was told to her that the screws were too long and there were no end caps on the screws.
 
Manufacturer Narrative
Investigation results were made available.Event description: it was reported that patient had a distal femur fracture which was treated with ncba periprosthetic femur plate.Post-implantation the patient started experiencing pain, extreme swelling around her knee and limited range of motion.Review of received data: no medical data relevant to the case has been received.X-rays: images were provided which are a xeroxed copy of actual x-ray and are grainy and of poor quality.Therefore, an assessment was not possible.Product evaluation: no product was returned as it remains implanted; therefore, visual and dimensional evaluation could not be performed.Review of product documentation: device purpose: all involved devices are intended for treatment.Product compatibility: the product combination was approved by zimmer biomet.Dhr review: review of the device history records could not be performed as the lot numbers of the devices were not available.Instruction for use (ifu): the applicable instruction for use (ifu) was reviewed with no anomalies noted.Conclusion: it was reported that patient had a distal femur fracture which was treated with ncba periprosthetic femur plate.Post-implantation the patient started experiencing pain, extreme swelling around her knee and limited range of motion.Neither operative notes, office visit notes, nor device(s) or photos of the device(s) were received; therefore the condition of the component(s) is unknown.Patient factors that may have affected the performance of the components such as bone quality, activity level, type of activity (low impact vs.High impact), and relevant medical history are unknown.Adherence to rehabilitation protocol is unknown.Based on the investigation the reported event cannot be confirmed.Due to significant lack of information a detailed investigation could not be performed, nevertheless based on the given information there is no indication of a nonconformance or complaint out of box (coob).Based on the given information and the results of the investigation, we were not able to identify a specific root cause for this issue.The need for corrective measures is not indicated and zimmer switzerland manufacturing gmbh considers this case as closed.Zimmer biomet's reference number of this file is (b)(4).The following reports are associated with this event: 0009613350-2021-00043-1; 0009613350-2021-00040-1; 0009613350-2021-00046-1; 0009613350-2021-00044-1; 0009613350-2021-00041-1; 0009613350-2021-00047-1; 0009613350-2021-00042-1.
 
Event Description
Investigation has been completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCB, SCREW, 5.0, 34 MM
Type of Device
NCB PLATING SYSTEM
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
MDR Report Key11242742
MDR Text Key229085805
Report Number0009613350-2021-00045
Device Sequence Number1
Product Code HRS
UDI-Device Identifier00889024295735
UDI-Public00889024295735
Combination Product (y/n)N
PMA/PMN Number
K042695
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number02.03150.034
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Weight113
-
-