Model Number N/A |
Device Problem
Inability to Auto-Fill (1044)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp and was unable to reproduce the reported "autofill failure" issue, although one incident was found in the fault logs.The iabp was able to pass all the leak tests, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) had an autofill failure.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|
Manufacturer Narrative
|
Corrected fields: e1 (event site email).
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) had an autofill failure.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) had an autofill failure.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g3, g6, h2, h6, h10.
|
|
Search Alerts/Recalls
|