Device evaluated by mfr: synergy ous mr 2.50 x 28 mm stent delivery system was returned for analysis.A visual examination of the stent found evidence of damage.Struts on the proximal end were lifted and bunched.The undamaged crimped stent outer diameter was measured using snap gauge and the result was 0.0385, this is within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found no issues.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.A 0.014 test guidewire loaded successfully.No other issues were identified during the product analysis.
|