• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC L3W0750 - DUODERM; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC L3W0750 - DUODERM; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 187957
Device Problems Delivered as Unsterile Product (1421); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device 1 of 2.Initial reporter: (b)(6).Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Event Description
It was reported that there was foreign body (paper) found on the dressing.The product was not used on patient.A photograph depicting the issue was received from the complainant.
 
Event Description
To date no additional patient or event details have been received.
 
Manufacturer Narrative
Correction (g1) - contact office address: dr.(b)(6).Batch record review: lot: 9m01493 was manufactured on 12/11/2019 in the bodolay line with a total of (b)(4) market units.Complaint investigator id: (b)(4) performed a batch record review on 04/30/2021 to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bom and all the tooling information documented was also correct, under icc code 187957 sap material id: (b)(4) and manufacturing order: (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.Returned sample evaluation: photo related to the reported problem is available for evaluation.Investigation summary: based in the analysis phase conclusions.For extra material in seal or foreign matter the root causes found were: manpower: incorrect dressing inspection: residues of paper and film stick to dressing after elc (extrusion lamination and cutting) process, since the scrap collector is above the dressing web.Since dressings are automatically feed in the bodolay machine, the units are packaged without the operator and vision system noticing.An opportunity was found related to the inspection process performed by the elc 10 and elc 11 since dressings must be inspected 100% as per applicable pi31-103 ver 17.0 and pi31-141 ver 7.0.For dirty on package (black stain) the root causes found were: method: unclear steps to perform online rework: an opportunity was found since there is not a standardized method for the segregation and/or rework of the blisters after having a failure or breakdown on the cartoner machine.Wrong placement of blisters (method).During the indexation process of blisters from the primary packaging to secondary packaging, not all blisters are not organized into the counter pockets.It is a manual process.This issue causes that the blisters do not enter on the mku¿s, creating defects such as torn or crushed or dirty package.Should additional information become available, a follow up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 9618003.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L3W0750 - DUODERM
Type of Device
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
MDR Report Key11250321
MDR Text Key229324783
Report Number9618003-2021-00164
Device Sequence Number1
Product Code NAD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number187957
Device Lot Number9M01493
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-