MAUDE Adverse Event Report: CORDIS CORPORATION UNKNOWN OPTEASE VENA CAVA FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION UNKNOWN OPTEASE VENA CAVA FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number 466FXXXX
Device Problems Migration or Expulsion of Device (1395); Difficult to Remove (1528); Material Twisted/Bent (2981); Unintended Movement (3026)
Patient Problems Internal Organ Perforation (1987); Perforation of Vessels (2135); Stenosis (2263); Device Embedded In Tissue or Plaque (3165)
Event Date 10/30/2020
Event Type  Injury  
Manufacturer Narrative

As reported, a patient underwent placement of an optease inferior vena cava (ivc) filter. The indication for filter placement was not reported. The filter subsequently malfunctioned including, but not limited to embedment, perforation of filter struts beyond the wall of the ivc, perforation into surrounding organs/tissues and ivc stenosis. The filter remains implanted; thus, unavailable for analysis. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed. The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. The predominant concern for embedding with in the wall of the ivc is the development of endothelialization. Endothelialization is the the healing of the inner surfaces of vessels or grafts by endothelial cells. This is the normal process whereby the body heals and recovers from invasive procedures. Endothelialization has been shown to occur in as short a period as 12 days. It was reported that there was perforation of the ivc and organs; however, a clinical conclusion could not be determined as to the cause of the event. A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters. Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place. Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation. Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics. Without procedural films or images for review the reported event(s) could not be confirmed. Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.

 
Event Description

As reported in the legal brief, a patient underwent placement of an optease vena cava filter. The filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to embedment, perforation of filter struts beyond the wall of the inferior vena cava (ivc), perforation into surrounding organs/tissues and ivc stenosis. As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages and required extensive medical care and treatment. As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering and other damages.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN OPTEASE VENA CAVA FILTER
Type of DeviceFILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014 2802
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014 2802
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014-2802
7863138372
MDR Report Key11250364
MDR Text Key229300527
Report Number1016427-2021-04742
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK034050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/22/2021
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received01/29/2021
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number466FXXXX
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/26/2021
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 01/29/2021 Patient Sequence Number: 1
-
-