SIRONA DENTAL SYSTEMS GMBH MIDWEST TRADITION HANDPIECE HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 780344
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional serial numbers included in this report: (b)(4). 9 of 10 devices were returned for evaluation. 1 device will not be returned for evaluation. Evaluation of 4 devices found normal chuck wear and the turbines needed to be replaced. The devices were repaired and returned to the customers. Evaluation of 5 devices found chuck wear and lack of maintenance. The devices were cleaned of debris and the turbines were replaced. The devices were repaired and returned to the customers.
 
Event Description
This report summarizes 10 malfunction events where midwest tradition handpieces would not hold burs. No injury resulted in any of the events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMIDWEST TRADITION HANDPIECE
Type of DeviceHANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM D-64625
Manufacturer Contact
karl nittinger
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494424
MDR Report Key11256682
MDR Text Key229595872
Report Number9614977-2020-00055
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K131319
Number of Events Reported10
Summary Report (Y/N)Y
Report Source Manufacturer
Type of Report Initial
Report Date 01/31/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/31/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator
Device Catalogue Number780344
Was Device Available for Evaluation? No Answer Provided
Was the Report Sent to FDA?
Event Location No Information
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? No Answer Provided
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage

-
-