• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 190713
Device Problems Thermal Decomposition of Device (1071); Smoking (1585); Sparking (2595)
Patient Problem No Patient Involvement (2645)
Event Date 01/11/2021
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation. Additionally, no on-site evaluation was performed by a fresenius field service technician (fst). A records review was performed on the reported serial number. An investigation of the device manufacturing records was conducted by the manufacturer. There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event. In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements. The investigation into the cause of the reported problem was not able to be confirmed. A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Event Description
A user facility biomedical technician (biomed) contacted fresenius technical support (ts) to report that a 2008t hemodialysis (hd) machine sparked and then smoked during a chemical disinfect cycle. The biomed traced the issue to the actuator board and the bibag interface board. Both parts displayed visible burn damage, but the actuator board looked particularly damaged. The biomed stated that every chip on the actuator board was pitted. The biomed also noticed a burning smell. As the event played out, the display on the machine went blank. The biomed unplugged the machine. The smoke detector did not go off and a fire extinguisher was not needed. The biomed reported that the smoking was minimal. They replaced the actuator board, the bibag interface board, the power supply, and the power board. However, the display issues persisted. The biomed called ts back to further troubleshoot the issue. The biomed was advised to order a new functional board and/or user interface (ui) board. The biomed replaced these parts and the display issue was resolved. The machine was plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet, and there was no previous history of the machine failing the electrical leakage test. The biomed stated there were approximately 6,000 hours on the machine at the time of the event. After the repair was completed, the biomed sated the machine was returned to service. The replaced parts were not available to be returned for evaluation as they were reportedly discarded. Photographs of the burnt cable were also not available. There was no patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008T HEMODIALYSIS SYS., WITH CDX
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key11257775
MDR Text Key229577915
Report Number2937457-2021-00160
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100897
UDI-Public00840861100897
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/01/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number190713
Device Catalogue Number190713
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device AgeMO
Event Location No Information
Date Manufacturer Received01/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/10/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-