The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.Na.
|
It was reported that the procedure was performed to treat a de novo lesion in the heavily calcified, heavily tortuous, 90% stenosed right coronary artery.A 3.25x12mm nc traveler balloon dilatation catheter (bdc) experienced resistance with the anatomy, and the balloon ruptured during the third inflation at 12 atmospheres.An unspecified balloon was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|