• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE CARDIO PICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE CARDIO PICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS Back to Search Results
Model Number MERGE CARDIO 10.2
Device Problem Loss of Data (2903)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2021
Event Type  malfunction  
Event Description
Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information from merge. Healthcare and other vendors systems including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology. Signal (waveform) data. Merge cardio is intended to allow users to review diagnostic and non-diagnostic quality images, annotate. Studies, perform digital subtraction on images, to perform quantitative measurements on images (including but not limited to quantitative coronary analysis, left ventricular analysis, time, area, length, velocity, angle, volume, and velocity-time integrals), to generate physician generated clinical reports (via structure reporting and template based tools), and to store this information in a database. On (b)(6) 2021, a customer contacted merge healthcare to report that a recent report was not available for viewing. Merge healthcare support found that the report was never confirmed but was in confirmed status. The logs had been truncated so they were unavailable for troubleshooting purposes. Support manually reset the status of the report and it was then able to be read and confirmed by the physician. No further action is necessary at this time. This has the potential to delay patient treatment and/or diagnosis. There have been no reports of patient injury or harm as a result of this issue. Reference complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERGE CARDIO
Type of DevicePICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
brian bell
75 binney street
cambridge, MA 02142-1123
3123540189
MDR Report Key11263610
MDR Text Key257205763
Report Number2183926-2021-00001
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial
Report Date 02/01/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMERGE CARDIO 10.2
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-