The patient was noted to have an acute change in mental status and new left-sided weakness.Code stroke initiated, patient connected to zoll r series defibrillator with pads for continuous monitoring.The patient was transported to ct (computed tomography) with the defibrillator.The patient returned to the room following cta (computed tomography angiography) and the defibrillator was plugged into the wall.The monitor screen of the defibrillator began flashing a battery fault message although the device was plugged into an electrical outlet.The defibrillator was unplugged and plugged into a different electrical outlet, the battery fault message continued.Troubleshooting continued by activating a defibrillator check and the patient was inadvertently shocked with 35.8 joules.The patient suffered a cardiac arrest as a result of the shock.Cpr was immediately initiated.The patient was intubated and there was return of circulation after three rounds of cpr.The patient was transferred to icu.Suggest the manufacturer update equipment to advise the operator if a shock is not advised based on the rhythms.Manufacturer response for defibrillator, zoll r series defibrillator (per site reporter).Zoll replaced the internal battery, ecg isolator, ecg shields, and lcd cable.The device passed all final system-level tests.
|