Pt having dialysis run in.
In pt dialysis, lost pulse and code blue called.
Pt coded intubated and transferred to icu.
Manufacturer response for hemodialysis unit, gambro (per site reporter): when the unit came down to the shop.
I ran a setup with a pa and puf auto cal.
Performed a patient simulation for 30 minutes.
Unit passed.
Total amount removed according to unit: 800 ml-- measured amount 807.
9 ml (within 25ml +/-).
While the patient simulation passed without any errors/alarms.
I found a leak on the bic conductivity.
The leak was approximately 3 drops a minute.
Once the patient simulation was completed.
I proceeded to go through all of the tests that are required for an annual inspection.
The unit passed every test with no discrepancies being noted.
Once complete i discussed all of my findings/readings of the unit with baxter tech support.
We had discussed having baxter tech support come to the hospital to possibly inspect the unit.
But because i am certified on the unit and did not believe the unit was the cause of the incident and because the patient was only on the unit for a total of 4 minutes and passed all operational tests they stated that they did not need to come out.
After getting the go ahead from baxter, i replaced both the ace and bic conductivity cells and also the safety valve.
After all the components were installed.
I ran a setup with a pa and puf auto cal.
Performed a patient simulation for 30 minutes.
Unit passed.
Total amount removed according to unit: 800 ml-- measured amount 803.
4 ml (within 25ml +/-) completed a bleach/rinse.
Unit returned to service.
|