Model Number NX1000-3-A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Death (1802)
Event Date 01/05/2021
Event Type  Death  
Manufacturer Narrative
The device was received for evaluation and successfully passed the prime and alarms test confirming proper functionality of the device and sensors.Available log files were retrieved and analyzed which showed device performance was without incident and was unremarkable.A review of the device history record confirmed the product was released meeting all product design specifications and quality criteria.All information supports the device was functioning as designed and intended with no malfunction.The nxstage system one user guide states a trained and qualified person must respond promptly and appropriately to all alarms and harmful conditions during treatment.The nxstage system one user guide supplement for solo home hemodialysis outlines risks, warnings, and requirements for the solo home hemodialysis patient.(b)(4).
 
Event Description
A report was received on (b)(6) 2021 from the home therapy nurse (htn) of a (b)(6) male patient with a medical history of end stage renal disease approved for performing solo home hemodialysis therapy, stating the patient expired during a hemodialysis session on (b)(6) 2021.Although requested, additional information has not been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key11273837
MDR Text Key230097954
Report Number3003464075-2021-00004
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K122051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberNX1000-3-A
Device Catalogue NumberCHRONIC HI-FLOW CYCLER, ROHS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2021
Is the Reporter a Health Professional? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age52 YR
Patient Weight62
-
-