Model Number 50000000 |
Device Problem
Device Sensing Problem (2917)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
|
Event Description
|
It was reported that the arctic sun device got an alert 113 (reduced water temperature control) and was not calibrating.A functional check showed that the mixing pump had failed.They provided part no.And price for mixing pump device will be repaired onsite.
|
|
Event Description
|
It was reported that the arctic sun device got an alert 113 (reduced water temperature control) and was not calibrating.A functional check showed that the mixing pump had failed.They provided part no.And price for mixing pump device will be repaired onsite.
|
|
Manufacturer Narrative
|
Upon further review, bd has determined that this mdr was reported in error as it was found to be a cascading failure of an event previously reported.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
|
|