Model Number 9550
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event: used the first date of the month of the aware date as no event date was provided.Device evaluated by mfr.: promus premier ous mr 20 x 4.00 stent delivery system was returned for analysis.An examination (visual and via scope) of the crimped stent identified stent damage.Stent struts in the proximal region of the stent lifted and pulled distally.The stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon was reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found multiple hypotube kinks.A visual and tactile examination of the shaft polymer extrusion found no issues.An examination (visual and via scope) found no issues with the tip.No other device issues noted.
 
Event Description
Reportable based on device analysis completed on 14-jan-2021 it was reported that device damage occurred.A 20 x 4.00mm promus premier drug eluting stent was selected for use.However, it was noted that the material was defective.No patient complications were reported.However, returned device analysis revealed stent damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11274242
MDR Text Key230108407
Report Number2134265-2021-00545
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9550
Device Catalogue Number9550
Device Lot Number0021668965
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2020
Date Manufacturer Received01/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-