BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7015
Device Problems Material Deformation (2976); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter. The shaft, hypotube, tip and balloon were microscopically and visually examined. There were numerous kinks. There was a complete separation at 51. 6cm distal of the strain relief. There was contrast and blood in the inflation lumen, and blood in the guidewire lumen. The balloon was tightly folded. Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
Reportable based on device analysis completed on 14jan2021 it was reported that shaft kink was encountered. The 95% stenosed, target lesion was located in the mid left anterior descending artery. A 2. 75mm x 15mm quantum maverick balloon catheter was advanced for dilatation. However, it was noticed that the delivery shaft was kinked. The procedure was completed with another of the same device. There were no patient complications nor injuries reported and the patient's status was stable. However, returned device analysis revealed shaft detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameQUANTUM MAVERICK
Type of DeviceCATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11282883
MDR Text Key230470043
Report Number2134265-2021-01306
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P860019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial
Report Date 02/04/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number7015
Device Catalogue Number7015
Device Lot Number0024903859
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-