Model Number 1012276-15
Device Problems Deflation Problem (1149); Difficult to Remove (1528); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/24/2020
Event Type  Injury  
Manufacturer Narrative
The device was not returned for evaluation. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue. The investigation was unable to determine a conclusive cause for the reported deflation issue; however, the reported difficulty removing the device, balloon separation and additional treatment appear to be related to circumstances of the procedure. There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that this was a procedure performed to treat a lesion in the right coronary artery (rca) with mild calcification and no tortuosity. During the procedure, a 3. 5 x 15 mm rx trek balloon catheter was used for post dilatation. The balloon was inflated to nominal pressure. The balloon had difficulty deflating. During the attempt to remove the balloon catheter, there was resistance due to the balloon not being fully deflated and the balloon separated. A snare device was used to remove separated balloon successfully. Outside the patient anatomy, it was noted there was still some contrast in the balloon. The procedure was completed. There was no adverse patient sequela. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTREK CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key11289308
MDR Text Key230676509
Report Number2024168-2021-00925
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 02/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1012276-15
Device Catalogue Number1012276-15
Device Lot Number00208G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/05/2021 Patient Sequence Number: 1
-
-