Model Number CLK-4
Device Problem Poor Quality Image (1408)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user facility that during an unspecified procedure with the subject device at the user facility, it was found that the light from the subject device was so dim that the tissue could not be clearly observed.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus.As the serial number was not provided, so olympus medical systems corp.Could not review the manufacturing history (dhr) of the subject device.The exact cause of the reported phenomenon could not be conclusively determined.However, we presume that the reported phenomenon was attributed to unsuitable lamp installation, aging deterioration of the lamp, inadequate setting of the brightness of the subject device, or the unstable connection of the endoscope.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11290408
MDR Text Key240417447
Report Number8010047-2021-02312
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170024528
UDI-Public04953170024528
Combination Product (y/n)N
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCLK-4
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-