It was reported that a pediatric patient became hypotensive and experienced tachycardia approximately two hours after starting continuous renal replacement therapy using a prismaflex control unit and a prismaflex m60 filter.
The patient was treated with a fluid bolus and one dose of calcium.
The patient's condition was reported to have stabilized.
On the same day, reported as later in the evening, a check return alarm, immediately followed by a return disconnection alarm and then immediately followed by an air in blood alarm was generated by the prismaflex control unit.
No disconnection or air in the return line was observed.
The air in blood alarm could not be cleared and the nurse was unable to hit stop during the air in blood alarm.
The machine was turned off and the blood was not returned.
The amount of blood loss was not reported.
The patient was re-connected to a new prismaflex control unit to continue with treatment.
No additional information is available.
|