SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72200616 |
Device Problems
Mechanical Jam (2983); No Apparent Adverse Event (3189)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.
There was a relationship found between the returned device and the reported incident.
A visual inspection was performed on the product and observed no issues.
A functional evaluation revealed a jammed motor/blade stall.
A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
A complaint history review concluded this was a repeat issue.
The complaint was confirmed and the root cause has been associated with a mechanical component failure.
Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.
No containment or corrective actions are recommended at this time.
Reference: case-(b)(4).
|
|
Event Description
|
It was reported that the mdu was not working.
It is unknown if there was a delay.
It is unknown whether the event happened during surgery and if there was patient involvement.
No further information.
Results of investigation have concluded that during functional evaluation, the device presented jammed motor/blade stall error, which makes it a reportable event.
All available information has been disclosed.
If additional information should become available, a supplemental report will be submitted accordingly.
|
|