The ventilator was investigated at site by our field service engineer.The ventilator alarmed for low o2 concentration and failed flow transducer test during pre-use check.The ventilator was making noise.The investigation found defective o2 gas module.The problem was solved by replicating the o2 gas module.After replacement, the ventilator passed all functional, safety, and electrical tests to factory specification.The ventilator was cleared for clinical use and returned to customer.No device logs were received and the replaced o2 gas module was not returned for investigation, therefor the root cause for the reported problem could not be determined.
|