Catalog Number 199725630S
Device Problems Entrapment of Device (1212); Material Twisted/Bent (2981)
Patient Problems Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2021
Event Type  Malfunction  
Manufacturer Narrative

Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received. Without a lot number the device history records review could not be completed. Product was not returned. Based on the information available, it has been determined that no corrective and/or preventative action is proposed. This complaint will be accounted for and monitored via post market surveillance activities. If additional information is made available, the investigation will be updated as applicable. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.

 
Event Description

Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2021 during a posterior lumbar interbody fusion (plif) treating spondylolisthesis at l5/s, after a rod was deployed, the collection key was not able to be tightened. Also, metal powders came out. Two (2) additional collection keys were tried but both failed. The screw was replaced and all of the devices were tightened. The procedure was delayed less than thirty (30) minutes. This report is for one (1) 5. 5 exp verse cannulated screw 6. 0 x 30. This is report 1 of 5 for (b)(4).

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name5.5 EXP VERSE CANNULATED SCREW 6.0 X 30
Type of DeviceORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ 02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key11301561
MDR Text Key230991812
Report Number1526439-2021-00261
Device Sequence Number1
Product Code NKB
Combination Product (Y/N)N
Reporter Country CodeJA
PMA/PMN NumberK142185
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 01/20/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received02/09/2021
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number199725630S
Device LOT Number209724
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer02/17/2021
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/01/2021
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured09/06/2018
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Unkown

-
-