The device was not returned for evaluation.
A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.
Based on the information reviewed, there is no indication of a product quality issue.
The reported patient effect of intimal dissection is listed in the xience xpedition everolimus eluting coronary stent system, instructions for use as a known patient effect of coronary stenting procedures.
A conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined; however, the subsequent treatment appears to be related to the operational context of the procedure.
There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
|
It was reported that the procedure was performed to treat a 90% stenosed, heavily tortuous, and moderately calcified de novo lesion in the right coronary artery.
The patient presented with angina.
A 3.
5x38mm xience xpedition stent was deployed; however, post-stenting an edge dissection was noted.
The dissection was successfully covered with an unspecified xpedition stent.
There were no adverse patient sequela and no clinically significant delay in the procedure.
No additional information was provided.
|