Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2021
Event Type  Injury  
Event Description
It was reported that stent profile issue occurred.The 70% stenosed target lesion was located in the mildly tortuous and moderately calcified mid right coronary artery.A 3.0x16mm synergy drug-eluting stent was advanced to treat the lesion.However, after stent deployment, it was noted that something on the distal stent needs to be covered.Another 3.0x16mm synergy stent was advanced through the first stent but failed to be delivered.The second stent was removed, then a picture was taken, and something was noted in the first stent.The balloon was checked and the stent was not there.The second synergy stent was stripped off and inside the first stent.A balloon was inserted through the stent and deployed it inside and proximal to the first stent.The procedure was completed with no further complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11306870
MDR Text Key231124196
Report Number2134265-2021-01030
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-