Model Number IPN001112 |
Device Problem
Incomplete or Inadequate Connection (4037)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
There was no patient involvement.It was reported that a purge failure occurred during preventative maintenance (pm) on the intra-aortic balloon pump (iabp).As a result, the pcs was replaced.
|
|
Manufacturer Narrative
|
Qn#(b)(4).Teleflex received the device for investigation.The reported complaint of purge failure alarm is confirmed.The returned pcs assembly's balloon pressure transducer zero offset was noted to be above specification during the complaint investigation, which caused the alarm.The root cause of this complaint is undetermined.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.A non-conformance has been initiated to further investigate the issue.
|
|
Event Description
|
There was no patient involvement.It was reported that a purge failure occurred during preventative maintenance (pm) on the intra-aortic balloon pump (iabp).As a result, the pcs was replaced.
|
|
Search Alerts/Recalls
|