Catalog Number 2N3328
Device Problem Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Additional initial reporter phone no.: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unspecified quantity of interlink system non-dehp t-connector extension sets have a defective t-connector.The defect was further described as the male luer-lock portion of the t-connector may be broken or has a ¿melted¿ appearance.This was identified during device inspection prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph which observed that the non-retractable male luer lock was damaged.The reported condition was verified.The cause of the condition was due to a machine issue during the manufacturing process.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK SYSTEM T-CONNECTOR EXTENSION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11309544
MDR Text Key232456499
Report Number1416980-2021-00530
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412004662
UDI-Public(01)00085412004662
Combination Product (y/n)Y
PMA/PMN Number
K921899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2N3328
Device Lot NumberUR20J11052
Was Device Available for Evaluation? No
Date Manufacturer Received03/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-