• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RFB MDU HAND CNTRL PWRMX EL SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RFB MDU HAND CNTRL PWRMX EL SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616F
Device Problems Mechanical Jam (2983); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2020
Event Type  malfunction  
Manufacturer Narrative
The reported device, used in treatment, was received for evaluation. There was a relationship found between the returned device and the reported incident. A visual inspection was performed on the product and observed no issues. A functional evaluation revealed a jammed motor/ blade stall. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended. A complaint history review concluded this was a repeat issue. The complaint was confirmed and the root cause has been associated with a mechanical component failure. Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out. No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that, during a shoulder arthroscopy, the "mdu powermax elite shaver" did not work. The procedure was successfully completed without significant delay using a back-up device. No patient injuries or other complications were reported. Results of investigation have concluded that during a functional evaluation, the device presented jammed motor/blade stall error, which makes it a reportable event. All available information has been disclosed. If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameRFB MDU HAND CNTRL PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin
5123913905
MDR Report Key11313222
MDR Text Key231743553
Report Number1643264-2021-00592
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/11/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number72200616F
Device Catalogue Number72200616F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-