Model Number 10-85-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(6) received a report that a s5 double head pump was giving a motor control failure error message during procedure.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H10: a livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to a defective pump head.The part was replaced and problem was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
|
|
Event Description
|
See initial report.
|
|