BAXTER HEALTHCARE CORPORATION CLEARLINK SOLUTION ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8858
Device Problem Protective Measures Problem (3015)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/14/2021
Event Type  malfunction  
Manufacturer Narrative
Device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported two (2) paclitaxel sets were causing spectrum pumps to alarm.During monoclonal antibody infusions the pump is observed to alarm ¿upstream occlusion¿.The set up is correct and the tubing appears ¿too rigid for our iv pumps¿.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the device was not returned.And the lot number is unknown.Therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK SOLUTION ADMINISTRATION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11323915
MDR Text Key232090664
Report Number1416980-2021-00596
Device Sequence Number1
Product Code FPA
Combination Product (y/n)Y
PMA/PMN Number
K160007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8858
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received03/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MONOCLONAL ANTIBODIES; SPECTRUM INFUSION PUMP
-
-