MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES ANSPACH EMAX 2 PLUS HAND PIECE - ROHS; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES ANSPACH EMAX 2 PLUS HAND PIECE - ROHS; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSR101209
Device Problems Overheating of Device (1437); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/22/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference number: (b)(4).
 
Event Description
It was reported that during a navio preventive maintenance, it was found that one anspach drill gets hot but also rotates 360.No patient was involved.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the navio drill, part number 101209, s/n (b)(6), intended for use in treatment was returned for evaluation.A relationship between the reported event and the device was established.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A functional evaluation was performed.The reported problem was confirmed.Nothing was identified visually that contributed to the reported problem.The device overheated and produced a louder than expected noise.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event is mechanical component failure of the motor shaft.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANSPACH EMAX 2 PLUS HAND PIECE - ROHS
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11326318
MDR Text Key232322170
Report Number3010266064-2021-00104
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPFSR101209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2021
Date Manufacturer Received09/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NPFS02000/NAVIO SURGICAL SYSTEM US/SN (B)(6)
-
-