DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Break (1069); Power Problem (3010)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate.
The fse evaluated the iabp, but found no problem with the power cord nor found it to be damaged.
Unrelated to the complained issue at hand, the fse was also unable to reproduce the reported gas leak alarm.
Moreover, no leaks were found upon review of the unit's logs.
The fse performed all functional and safety checks to meet factory specifications.
The iabp was then released to the customer for return to clinical service.
(b)(6).
|
|
Event Description
|
It was reported that during patient use, the cardiosave intra-aortic balloon pump (iabp) was found to have a damaged power cord.
No patient harm, serious injury or adverse event was reported.
|
|