• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 8022 HANDPC TORNADO SHAVER LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 8022 HANDPC TORNADO SHAVER LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 288022
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2021
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4). Investigation summary: the complaint device was received at the service center and evaluated. It was reported that does not turn ¿ jammed motor. Per service reports, this complaint can be confirmed. It was found during evaluation that the motor was noisy, and cable was out of range. Further, locking system was rusty and screws were defective. Drill mounting mechanism was repaired and the motor and motor cable were replaced to resolve the issues. After repair, the device was found to be working according to the specifications. Fluid ingress into the device and contact with the locking system is responsible for the rusty locking system. However, with the available information, we cannot determine the root cause of the other identified failures. The defective drill mounting mechanism, defective motor and defective motor cable would have caused the customer to experience the reported problem. The service history has been reviewed in lieu of the device history record for this device since it was previously serviced. The device was last serviced on 06/19/2018 and passed all functional testing before being returned to the customer. At this point in time, no corrective action is required, and no further action is warranted. Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the affiliate in (b)(6) that during an arthroscopy procedure on (b)(6) 2021, it was observed that the 8022 handpc tornado shaver device had a jammed motor and did not turn. During in-house engineering evaluation, it was determined that the locking system on the device was corroded. Another like device was used to complete the procedure without delay. There were no adverse patient consequences reported. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name8022 HANDPC TORNADO SHAVER
Type of DeviceLINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key11334075
MDR Text Key247933045
Report Number1221934-2021-00536
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeLO
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/10/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number288022
Device Catalogue Number288022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/10/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-