Catalog Number UNKNOWN FILTER
Device Problems Migration or Expulsion of Device (1395); Patient-Device Incompatibility (2682)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 02/27/2020
Event Type  Injury  
Manufacturer Narrative
Manufacturing review: the device history record review could not be performed as the lot number is unknown. Investigation summary: the device was not returned for evaluation. Medical records were not provided. The investigation is inconclusive for the alleged perforation of the inferior vena cava (ivc) and filter migration as no objective evidence has been provided to confirm any alleged deficiency with the filter. Based upon the available information a definitive root cause could not be determined. Labeling review: a review of product labeling documentation (e. G. Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism. Approximately fourteen years post filter deployment, it was alleged that filter migrated, perforated. The device has been removed surgically. The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVENA CAVA FILTER
Type of DeviceVENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key11337185
MDR Text Key232174529
Report Number2020394-2021-00276
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation
Type of Report Initial
Report Date 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberUNKNOWN FILTER
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/17/2021 Patient Sequence Number: 1
-
-